Calibration
Procedure
—
Type
502A
Fig.
5-4. Adjustment
of the GEOM
control. In waveform A the GEOM control
is
improperly set causing bowing of the vertical lines. In
waveform
B the
bowing of the vertical
lines has been reduced to a minimum by the proper adjustment of the GEOM control.
vertical
portions
of
the markers are displayed on the face
of
the
crt and adjust the TRIGGERING
LEVEL Control to
obtain
a
stable
display. Set the
GEOM control for minimum
curvature
of the vertical
lines at the sides of the crt.
8. Adjusting CRT
HORIZ SENS
BAL and
HORIZ
BEAM
REGIS
From
the Type 180A apply 100 MICROSECONDS
and 1
MILLISECONDS
markers to the A input of the LOWER BEAM
and
the B input of the UPPER BEAM. Switch the UPPER
BEAM
Input
Selector to B-DC and the SENSITIVITY switches
of
both beams to 5 VOLTS PER CM. With the Vertical
POSITION Controls,
position the traces until the markers of
the
LOWER
BEAM are just touching the markers of
the
UPPER BEAM.
Adjust the
HORIZ BEAM REGIS until the
first and ninth
time markers are
as near being above one another as
possible.
Now adjust
the CRT HORIZ SENS BAL until the
sensitivity of
the two beams is the same. At this point in
the
calibration the two beams are being matched to each
other and
the
time markers may not coincide with the
graticule
marks.
NOTE
Every
time the CRT HORIZ SENS BAL is adjusted
the INT
BAL
must be checked. The HORIZ BEAM
REGIS
may
also have to be readjusted at this time.
The
above
three
controls interact with each other.
9.
X1 CAL
and
X20
CAL Adjustments
Set
the
UPPER BEAM
Input Selector
to the A-DC position.
With
1
MILLISECONDS and
100 MICROSECONDS time
markers
applied
to
the A input of each beam reset the
HORIZONTAL
DISPLAY switch to X20 position. Adjust
the
X20
CAL for a display of one 100 MICROSECONDS time
marker
per 2 cm.
Now turn the HORIZONTAL DISPLAY
Switch
to X1 and adjust the X1 CAL for one 1 MILLI
SECONDS
time marker every cm.
Fig.
5-5.
Adjustment of the X20 control. The X20 CAL control is
adjusted
in
the
calibration procedure
for the correct magnified
sweep
rate. The control should be set
so that markers occur at
2
centimeter intervals.
There
will
be
interaction between the X1 and X20 adjust
ments.
After this interaction has been removed
the X2 and
X5 and
X10 positions
of the HORIZONTAL DISPLAY switch
should
be checked. Disconnect the
time markers.
5-4
Summary of Contents for 502A
Page 4: ......
Page 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Page 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Page 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Page 34: ...NOTES ...
Page 56: ...NOTES ...
Page 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Page 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Page 67: ...Parts List Type 502A TOP 6 11 ...
Page 70: ...Parts List Type 502A BOTTOM 6 14 ...
Page 73: ...Parts List Type 502A REAR 6 17 ...
Page 89: ...TYPE 502A OSCILLOSCOPE A ...
Page 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Page 91: ... TYPE 502A OSCILLOSCOPE A ...
Page 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Page 93: ......
Page 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Page 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Page 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Page 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Page 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Page 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Page 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Page 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Page 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...