![Tektronix 502A Instruction Manual Download Page 11](http://html1.mh-extra.com/html/tektronix/502a/502a_instruction-manual_1078909011.webp)
Operating
Instructions
—Type
502A
gering
will
normally be used when you are
observing a
waveform
which bears a fixed time relationship to the power
line frequency.
4. To
trigger the sweep from some external waveform
bear
ing a
definite
time relationship to
the
observed waveform,
connect
the
external
waveform to the TRIGGER INPUT con
nector
and set the TRIGGER SELECTOR switch at EXT AC
or
EXT DC.
External
triggering provide definite
advantages over other
configurations
of triggering in certain cases. With external
triggering,
the
triggering signal generally remains essen
tially constant
in amplitude and shape. It is thereby possible
to
observe the shaping and
amplification of a signal
by each
stage
of a
circuit
without
resetting the triggering controls
for
each observation. Also time and phase relationships be
tween
the waveforms at different points in the circuit can be
seen.
If, for example,
the external triggering
signal is de
rived
from the
waveform
at the input to a circuit, the time
relationship
and phase of the waveforms at each point in
the circuit are automatically compared
to the
input signal by
the
display presented on the face of
the crt.
Automatic Mode
Automatic
triggering
is obtained
by rotating the TRIGGER
ING
LEVEL
control fully counterclockwise to the AUTO
MATIC
position.
This
provides a
preset triggering level which
is set
to allow triggering at the
average voltage point of the
applied waveform.
The sweep runs
at approximately a 50
cycle rate when
no triggering
signals are applied. Auto
matic
triggering
can be used
with triggering signals obtained
from the LINE,
UPPER, LOWER, or
EXT positions of the TRIG
GER
SELECTOR switch, but for most waveforms, it is useful
only
for
triggering at
frequencies above approximately 50
cycles.
Automatic triggering saves considerable time
in
observing
a
series
of waveforms since
it is not necessary to
reset
the triggering level
for each observation.
DC
Coupling
DC coupling
is selected in the UPPER DC, LOWER DC, and
EXT
DC
positions
of the TRIGGER SELECTOR switch. This
coupling
of the
triggering signal is particularly useful in
triggering
from waveforms
which
are not adaptable to ac
coupling,
such as random pulse trains or very low frequency
waveforms.
Random pulse trains pose a
special problem in
ac
coupling
since the random occurrence
of the input wave
forms
causes the
average voltage
level to shift. This in turn
may
cause the triggering level
to shift to
an unstable point.
This
problem is not encountered
with dc coupling
since the
triggering
point is determined
only by instantaneous voltages.
With
dc
coupling the triggering signal is obtained from
either
the LOWER
or the
UPPER
BEAM amplifiers, hence
varying
the
respective
vertical
position
controls will change
the
triggering
point. For this reason, you
may find it neces
sary
to readjust
the TRIGGERING
LEVEL controls when
you
change
the
vertical position of the
trace. If you desire to
eliminate
this
effect, you can use ac
coupling provided the
triggering
signal is otherwise suitable for this type of coupling.
With
dc coupling, the dc level of external triggering signals
will
also
affect the triggering point.
Generally,
when the
triggering
signal is small
compared to its
dc level, ac coup
ling
should be used.
AC
Coupling
AC
coupling is selected in
the LINE, LOWER AC, UPPER
AC, and
EXT AC
positions of the TRIGGER SELECTOR switch.
This
coupling
provides
stable triggering on virtually all types
of
waveforms.
As a general rule, however, ac
coupling is
unsatisfactory
for
triggering with low amplitude waveforms
at frequencies
below approximately
15 cycles. This figure
will
vary depending
upon
the amplitude
and shape of the
triggering waveform
and should not therefore be set as an
absolute
standard.
Triggering at
frequencies
below 15 cycles
can
be
accomplished
when higher amplitude triggering
signals
are
used.
With
ac
coupling, the triggering point depends on the
average
voltage
level of the triggering signals. If the
triggering
signals occur
at random, the average voltage
level
will vary, causing the triggering point to also vary.
This
shift
of
the
triggering
point may
be enough so that it
is
impossible
to
maintain a stable display. In such cases
you
should
use dc coupling.
Trigger
Slope
If
you wish to trigger the sweep on the rising (positive
slope)
portion of the triggering waveform, place the TRIGGER
SELECTOR switch
at
+. If you
wish to trigger on the falling
(negative
slope)
portion of the triggering waveform, place
the
TRIGGER SELECTOR switch at -.
Using the
TRIGGERING LEVEL Control
In
all positions except AUTOMATIC and
RECURRENT, the
TRIGGERING
LEVEL Control determines at which point on the
triggering signal the
sweep is triggered. Using
this
control,
the
sweep
can
be triggered at any point on the waveform so
long
as the slope of the waveform is great enough to provide
stable
triggering. With dc coupling, the
sweep cannot be
triggered
with any degree of stability at the top of a square
wave,
for
example,
the time that the voltage remains
constant
is comparatively
long. As
a result, the sweep
triggers
at random points
along the top of the square wave
producing
considerable
trace jitter.
You can use the same method to set
the TRIGGERING
LEVEL control
for either type
of coupling. After selecting
the triggering slope,
rotate
the TRIGGERING LEVEL control
fully
counterclockwise to the AUTOMATIC position. Then
rotate
the TRIGGERING LEVEL control clockwise until the
sweep
no
longer
triggers. Continue to rotate the control
in
the clockwise direction until the sweep
again triggers
and
a stable display is obtained. Further rotation of the
control
in the
clockwise direction causes the sweep to trigger
at more
positive
points on the
triggering waveform. See
Figure 2-8.
Free-Running
Operation
With
the Type 502A, you can get a periodic, free-running
sweep,
independent of any external triggering
or synchroniz
ing
signal by rotating the TRIGGERING LEVEL control
fully
clockwise
to the
RECURRENT position. This permits you to
observe
the UPPER and LOWER BEAM traces
without an in
put signal.
2-5
Summary of Contents for 502A
Page 4: ......
Page 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Page 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Page 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Page 34: ...NOTES ...
Page 56: ...NOTES ...
Page 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Page 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Page 67: ...Parts List Type 502A TOP 6 11 ...
Page 70: ...Parts List Type 502A BOTTOM 6 14 ...
Page 73: ...Parts List Type 502A REAR 6 17 ...
Page 89: ...TYPE 502A OSCILLOSCOPE A ...
Page 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Page 91: ... TYPE 502A OSCILLOSCOPE A ...
Page 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Page 93: ......
Page 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Page 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Page 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Page 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Page 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Page 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Page 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Page 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Page 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...