Operating
Instructions—Type
502A
Fig.
2-6.
Location of locking sleeve and knurled parts on the Type P6006 Probe.
the
probe body until a flat
top is obtained on the square
wave
waveform.
The locking sleeve
should now be tight
ened. After
the
locking sleeve has been tightened, the probe
compensation should be
checked once more to insure that
the
compensation did not change when the locking sleeve
was
tightened.
Triggered
Operation
In
order
to obtain a stable display of some recurrent
waveform,
it
will be necessary for you to trigger the hori
zontal
sweep at
the correct instant of time.
This will in
sure
that the sweep will start at the same point on the ob
served
waveform for
each cycle of operation. The sweep
should
be
triggered either by some waveform bearing a
fixed time
relationship
to the observed waveform or by the
observed waveform
itself. Either method will produce the
desired
stable display. If the waveform which is applied to
the
UPPER BEAM and the waveform which
is applied to the
LOWER
BEAM have a definite time relationship
to each
other
it is possible to trigger from either beam and obtain
simultaneously
a stable display of both waveforms.
The
following
instructions tell you how to select the
proper
triggering
signal for various applications of your
oscilloscope.
These
instructions
also
provide information
about
the advantages and
limitations of each triggering
con
figuration.
You
should attempt to become thoroughly familiar
with all
of the various triggering configurations in order
that
you may obtain maximum use from your instrument.
A
thorough knowledge
of the triggering configurations will
allow
you a greater selection of triggering
methods when
you are
confronted with a definite triggering problem.
How
to select the triggering signal source
1.
To trigger
the sweep from the waveform
displayed on the
UPPER
BEAM, set the
TRIGGER SELECTOR switch at UPPER
AC
or
UPPER DC.
2.
To
trigger the sweep from the waveform displayed on the
LOWER
BEAM, set the TRIGGER SELECTOR switch at LOWER
AC or
LOWER DC.
3. To trigger the sweep at the power
line frequency, set
the
TRIGGER
SELECTOR switch at LINE. This mode of trig
Fig.
2-7. The Type P6006 Probe is adjusted for a flat-topped square-wave display of the Calibrator waveform.
2-4
Summary of Contents for 502A
Page 4: ......
Page 22: ...Circuit Description Type 502A Fig 3 2 Simplified Sweep Trigger Circuit t i CO ...
Page 27: ...Circuit Description Type 502A Fig 3 4 Simplified Horizontal Amplifier 3 9 ...
Page 32: ...Circuit Description Type 5O2A Fig 3 7 Simplified Calibrator Circuit 3 14 ...
Page 34: ...NOTES ...
Page 56: ...NOTES ...
Page 60: ...Parts List Type 502A LEFT SIDE 6 4 ...
Page 64: ...Parts List Type 502A RIGHT SIDE 6 8 ...
Page 67: ...Parts List Type 502A TOP 6 11 ...
Page 70: ...Parts List Type 502A BOTTOM 6 14 ...
Page 73: ...Parts List Type 502A REAR 6 17 ...
Page 89: ...TYPE 502A OSCILLOSCOPE A ...
Page 90: ...A BLOCK DIAGRAM MRH Z6 3 ...
Page 91: ... TYPE 502A OSCILLOSCOPE A ...
Page 92: ...A CIRCUIT NUMBERS 1 THRU 59 I Ixj 263 TIME BASE TR IGGER ...
Page 93: ......
Page 95: ...TI M I NG RESI 5TORS TIMING CAPACITORS TYPE 502A OSCILLOSCOPE ...
Page 96: ...SWI6O HOLD OFF CAPACITORS 4 RESISTORS TIM ING CAPACITORS o A TIMING SWITCH job ...
Page 98: ... 1 INPUT AMPLIFIER OUTPUT AMPLIFIER A CIR CUIT NUMSER 5 300 THR U 399 ZG3 joe ...
Page 100: ...CMO 363 VERTICAL ATTENUATOR SWITCH A ...
Page 103: ...TYPE 5O2A OSCILLOSCOPE OOM HEATER WIRING DIAGRAM ...
Page 104: ...POWER SUPPLY CIRCUIT NUMBERSI 600 THRU 799 ...
Page 106: ...CIR CUIT NUMBER S 800 THR U 869 CRT CIRCUIT c 9G4 ...
Page 107: ...CA LIBp A TOp MULTIVIBP A TOR 1 IOOV o TYPE 5O2A OSCILLOSCOPE A ...
Page 108: ...CAL OUT CF CALIBRATOR A CIRCUIT NUMBERS 8 7O THRU 899 4Z ...