Table 5
Electrical Characteristics
Characteristics
Performance Requirements
Supplemental Information
V E R TIC A L DEFLECTION SYSTEM
Deflection Factor
Range
5 mV per division to 5 V per division in
a 1, 2, 5 sequence.
Accu racy
±3% on all ranges when VO LTS/DIV is
calibrated at 5 mV per division; add
0.05% per °C deviation from 25°C.
Uncalibrated (VAR) Range
Continuously variable between VOLTS/
DIV switch settings. Reduces deflection
factor at least 2.5 to 1 on all VO LTS/DIV
switch settings.
Reduces deflection factor to at least
2 mV per division w ith VO LTS/DIV
switch set to 5 mV.
Frequency Response
6-division reference signal from a
25-£2 source; centered vertically, with
VO LTS/DIV VAR control in cali
brated detent.
— 15°C to +40°C
Dc to at least 100 MHz.
Reduces to 88 MHz at 2 mV per division.3
Perform ance Requirement not checked in Service Manual.
@