Time Difference Between Two Time-Related Pulses
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Using probes or cables having equal time delays,
apply the reference signal to the Channel 1 input and apply
the comparison signal to the Channel 2 input.
3. Set both VO LTS/D IV switches to produce a display
o f either 4 or 5 divisions in amplitude.
4. Select either A LT or CHOP VE R TIC A L MODE,
depending on the input frequencies (or select AUTO
V E R TIC A L MODE if automatic selection is desired).
5. Set the A SEC/DIV switch to display the measure
ment points o f interest w ithin the graticule area.
6. Select the A INTEN HORIZ MODE and set the
B SEC/DIV switch to the fastest sweep speed that provides
a usable (visible) intensified zone.
7. Adjust the B DELAY TIME POSITION control to
move the reference point to the reference signal (see
Figure 27A, Point A).
8. Adjust the A TIME POSITION control to move
the measurement point to the comparison signal (see
Figure 27A, Point B).
9. Press in the B HORIZ MODE push button and
adjust the B DELAY TIME POSITION and the A TIME
POSITION controls to superimpose the reference- and
comparison-signal leading edges as shown in Figure 27B.
10. Read the time difference on the LCD readout.
(P>
2237 Operators
65
Summary of Contents for 2337
Page 9: ... ...
Page 20: ...4119 06 Figure 5 Horizontal controls and indicator ...
Page 24: ...13 ire 6 A TRIGGER controls connector and indicator ...
Page 29: ...Figure 7 DMM and B TRIGGER controls connectors and LCD readout 16 2337 Operators ...
Page 72: ... Figure 22 Delayed sweep magnification ...
Page 81: ...Figure 27 Time difference between two time related pulses 66 2237 Operators ...
Page 116: ...NOTES ...
Page 117: ...NOTES ...