T a b le 5 (c o n t)
Characteristics
Performance Requirements
Supplemental Information
D IG IT A L M ULTIMETER (cont)
Dc Voltage (cont)
Rejection Ratio
Normal Mode
A t least 60 dB at 50 and 60 Hz.
Common Mode
A t least 100 dB at dc;
at least 60 dB at 50 and 60 Hz.
Response Time
No Autorange
W ithin 3 seconds.
Uprange
W ithin 9 seconds.
Downrange
W ithin 7 seconds.
Maximum Input Voltage
+ (positive) Input to Ground
500 V (dc + peak ac) at 60 Hz.a
— (negative) Input to Ground
500 V (dc + peak ac) at 60 Hz.a
Positive to Negative Inputs
500 V (dc + peak ac) at 60 Hz.a
Perform ance Requirement not checked in Service Manual.
@