(T?) BW L IM IT Switch—Limits the bandwidth o f the
vertical amplifier to approximately 20 MHz when
pressed in. Push button must be pressed a second
time to release it and regain fu ll 100-MHz bandwidth
operation. Provides a method fo r reducing inter
ference from unwanted high-frequency signals when
viewing low-frequency signals.
(fiT) TRIG VIEW Switch-Press in and hold this push
button to display a sample o f the signal present in
the A Trigger amplifier (for all A TRIGGER
SOURCE switch settings except VERT MODE).
A ll other signal displays are removed while the
TRIG VIEW push button is held in.
HORIZONTAL
Refer to Figure 5 for location o f items 20 through 26.
(20) B DELAY TIME POSITION-A TIME POSITION
Controls—Select the amount o f delay time between
start o f the A Sweep and start o f the B Sweep. Delay
time is variable to at least 10 times the A SEC/DIV
switch setting. The B DELAY TIME POSITION
(outer knob) controls the reference point when the