Time Difference Between Two Time-Related Pulses
The calibrated sweep speed and dual-trace features of
the 2237 allow measurement o f the time difference
between two separate events. To measure time difference,
use the following procedure:
1.
Preset instrument controls and obtain a Normal
Sweep Display.
2. Set the A TRIGGER SOURCE switch to CH 1.
3.
Set both AC-GND-DC switches to the same position,
depending on the type o f coupling desired.
4.
Using either probes or cables w ith equal time delays,
connect a known reference signal to the Channel 1 input
and the comparison signal to the Channel 2 input.
5.
Set both VO LTS/DIV switches fo r 4- or 5-division
displays.
6.
Select either A LT or CHOP VE R TIC A L MODE,
depending on the frequency of input signals (or select
AUTO VER TICAL MODE if automatic selection is desired).
7.
If the two signals are o f opposite polarity, press in
the CH 2 INVERT push button to invert the Channel 2
display (signals may be o f opposite polarity due to 180°
phase difference; if so, note this fo r use later in the final
calculation).
8.
Adjust the A TRIGGER LEVEL control fo r a stable
display.
9.
Set the A T IM E /D IV switch to a sweep speed which
provides three or more divisions o f horizontal separation
between the reference points on the two displays. Center
each o f the displays vertically (see Figure 19).
10.
Measure the horizontal difference between the two
signal reference points and calculate the time difference
using the following formula:
SEC/DIV
horizontal
switch
x
difference
Time
_
setting______ (divisions)
Difference
magnification factor
EXAMPLE:
The A SEC/DIV switch is set to 50 /is,
the X10 MAG switch is pressed in and the horizontal
50
2337 Operators
@
Summary of Contents for 2337
Page 9: ... ...
Page 20: ...4119 06 Figure 5 Horizontal controls and indicator ...
Page 24: ...13 ire 6 A TRIGGER controls connector and indicator ...
Page 29: ...Figure 7 DMM and B TRIGGER controls connectors and LCD readout 16 2337 Operators ...
Page 72: ... Figure 22 Delayed sweep magnification ...
Page 81: ...Figure 27 Time difference between two time related pulses 66 2237 Operators ...
Page 116: ...NOTES ...
Page 117: ...NOTES ...