T a b le 5 (c o n t)
Characteristics
Performance Requirements
Supplemental Information
TRIGGER SYSTEM (cont)
LEVEL Control Range (cont)
EXT -r 10
A t least ± 10 V, 20 V p-p.a
Trigger View (A Trigger)
Deflection Factor
EXT
100 mV per division ±40%.
EXT
t
10
1 V per division ±40%.
Centering o f Trigger Point
W ithin 1 division o f center screen.
Bandwidth
To at least 80 MHz.
4-division reference signal from a 25-J2
source; centered vertically.
Delay Difference
3 ns ±2 ns.
5-division signal with 5-ns rise time or
less from 25-S2 source, centered
vertically; equal cable length from
signal source to vertical channel and
external trigger inputs, terminated in
50
at each input.
Perform ance Requirement not checked in Service Manual.
@