Appendix D: Performance Tests
Appendices
AĆ64
d. Vertically center the display.
e. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 50 kHz.
f.
CHECK for a triggered display (
TRIG'D
indicator light on).
g. Set the
SEC/DIV
control to 50 ns.
h. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 10 MHz.
i.
CHECK for a triggered display (
TRIG'D
indicator light on).
j.
Disconnect the probe from the test equipment.
3. Check Motor Trigger
a. Set:CH 1
VOLTS/DIV
0.1 V
SEC/DIV
1 ms
Trigger
MODE
AUTO BL
Trigger
SOURCE
VERT
TRIG POS
MID
MOTOR TRIG
ON
b. Connect the fast rise output of the Calibration Generator
through the BNCĆtoĆbananaĆplug adapter to the CH 1 probe
tip and adjust the fast rise amplitude for 4 divisions.
c. Set the Calibration Generator frequency to 1 kHz.
d. Press the
AUTO:LVL PUSH
button to center the trigger level
on the waveform.
e. Adjust the Calibration Generator frequency variable to the
point where the trigger of the waveform is just able to be
maintained.
f.
Verify that the negative portion of the displayed square wave
is between 2.0 and 3.0 ms.
g. Disconnect the test equipment from the instrument.
Appendix D: Performance Tests
Appendices
AĆ64
d. Vertically center the display.
e. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 50 kHz.
f.
CHECK for a triggered display (
TRIG'D
indicator light on).
g. Set the
SEC/DIV
control to 50 ns.
h. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 10 MHz.
i.
CHECK for a triggered display (
TRIG'D
indicator light on).
j.
Disconnect the probe from the test equipment.
3. Check Motor Trigger
a. Set:CH 1
VOLTS/DIV
0.1 V
SEC/DIV
1 ms
Trigger
MODE
AUTO BL
Trigger
SOURCE
VERT
TRIG POS
MID
MOTOR TRIG
ON
b. Connect the fast rise output of the Calibration Generator
through the BNCĆtoĆbananaĆplug adapter to the CH 1 probe
tip and adjust the fast rise amplitude for 4 divisions.
c. Set the Calibration Generator frequency to 1 kHz.
d. Press the
AUTO:LVL PUSH
button to center the trigger level
on the waveform.
e. Adjust the Calibration Generator frequency variable to the
point where the trigger of the waveform is just able to be
maintained.
f.
Verify that the negative portion of the displayed square wave
is between 2.0 and 3.0 ms.
g. Disconnect the test equipment from the instrument.
Summary of Contents for 222PS
Page 6: ......
Page 18: ...Contents xii Contents xii ...
Page 21: ...Overview Overview ...
Page 22: ......
Page 27: ...At a Glance At a Glance ...
Page 28: ......
Page 45: ...In Detail In Detail ...
Page 46: ......
Page 50: ...Acquisition Modes In Detail 3Ć4 Acquisition Modes In Detail 3Ć4 ...
Page 54: ...Auto Setup In Detail 3Ć8 Auto Setup In Detail 3Ć8 ...
Page 60: ...Calibration In Detail 3Ć14 Calibration In Detail 3Ć14 ...
Page 66: ...Channels In Detail 3Ć20 Channels In Detail 3Ć20 ...
Page 70: ...The Display In Detail 3Ć24 The Display In Detail 3Ć24 ...
Page 76: ...Horizontal Operation In Detail 3Ć30 Horizontal Operation In Detail 3Ć30 ...
Page 96: ...Probes In Detail 3Ć50 Probes In Detail 3Ć50 ...
Page 106: ...Saving and Recalling Data In Detail 3Ć60 Saving and Recalling Data In Detail 3Ć60 ...
Page 108: ...Store Mode In Detail 3Ć62 Store Mode In Detail 3Ć62 ...
Page 124: ...Vertical Operation In Detail 3Ć78 Vertical Operation In Detail 3Ć78 ...
Page 127: ...Appendices Appendices ...
Page 128: ......
Page 170: ...Appendix C Quick Checks Appendices AĆ42 Appendix C Quick Checks Appendices AĆ42 ...
Page 210: ...Appendix F Accessories Appendices AĆ82 Appendix F Accessories Appendices AĆ82 ...
Page 211: ...Glossary Index Glossary Index ...
Page 212: ......
Page 230: ...Index Index IĆ10 Index Index IĆ10 ...
Page 231: ......
Page 232: ......