Appendix C: Quick Checks
222PS Operator Manual
AĆ39
Initial Setup Procedure
This procedure connects the oscilloscope to external power and
installs probes for the tests that follow.
Equipment Required (See Table AĆ8, page AĆ45)
P850 probe (standard accessory, Item 13)
External Power AC Adapter (standard accessory, Item 3)
Prerequisites
None
Procedure
1. Plug the External Power AC Adapter into the AC power source.
2. Plug in the cord from the adapter to the
EXT POWER INPUT
connector on the rear panel of the oscilloscope.
3. Open the zipper on the probe pouch and remove the probes. If
disconnected, connect the probes through the oval opening at the
rear of the pouch to the connectors on the oscilloscope.
4. Press the
ON
button of the oscilloscope to toggle it into the
operating mode.
Self Cal Tests
This procedure uses internal routines to verify that this 222PS functions
and can properly perform selfĆcalibration of both input channels. This
procedure also guarantees the highest accuracy state for the
PerformĆ
ance Tests
that follow.
Equipment Required (See Table AĆ8, page AĆ45)
Connector; femaleĆtoĆdualĆbananaĆplug (Item 7 )
External Power AC Adapter (standard accessory, Item 3)
Prerequisites
Initial Setup Procedure
Appendix C: Quick Checks
222PS Operator Manual
AĆ39
Initial Setup Procedure
This procedure connects the oscilloscope to external power and
installs probes for the tests that follow.
Equipment Required (See Table AĆ8, page AĆ45)
P850 probe (standard accessory, Item 13)
External Power AC Adapter (standard accessory, Item 3)
Prerequisites
None
Procedure
1. Plug the External Power AC Adapter into the AC power source.
2. Plug in the cord from the adapter to the
EXT POWER INPUT
connector on the rear panel of the oscilloscope.
3. Open the zipper on the probe pouch and remove the probes. If
disconnected, connect the probes through the oval opening at the
rear of the pouch to the connectors on the oscilloscope.
4. Press the
ON
button of the oscilloscope to toggle it into the
operating mode.
Self Cal Tests
This procedure uses internal routines to verify that this 222PS functions
and can properly perform selfĆcalibration of both input channels. This
procedure also guarantees the highest accuracy state for the
PerformĆ
ance Tests
that follow.
Equipment Required (See Table AĆ8, page AĆ45)
Connector; femaleĆtoĆdualĆbananaĆplug (Item 7 )
External Power AC Adapter (standard accessory, Item 3)
Prerequisites
Initial Setup Procedure
Summary of Contents for 222PS
Page 6: ......
Page 18: ...Contents xii Contents xii ...
Page 21: ...Overview Overview ...
Page 22: ......
Page 27: ...At a Glance At a Glance ...
Page 28: ......
Page 45: ...In Detail In Detail ...
Page 46: ......
Page 50: ...Acquisition Modes In Detail 3Ć4 Acquisition Modes In Detail 3Ć4 ...
Page 54: ...Auto Setup In Detail 3Ć8 Auto Setup In Detail 3Ć8 ...
Page 60: ...Calibration In Detail 3Ć14 Calibration In Detail 3Ć14 ...
Page 66: ...Channels In Detail 3Ć20 Channels In Detail 3Ć20 ...
Page 70: ...The Display In Detail 3Ć24 The Display In Detail 3Ć24 ...
Page 76: ...Horizontal Operation In Detail 3Ć30 Horizontal Operation In Detail 3Ć30 ...
Page 96: ...Probes In Detail 3Ć50 Probes In Detail 3Ć50 ...
Page 106: ...Saving and Recalling Data In Detail 3Ć60 Saving and Recalling Data In Detail 3Ć60 ...
Page 108: ...Store Mode In Detail 3Ć62 Store Mode In Detail 3Ć62 ...
Page 124: ...Vertical Operation In Detail 3Ć78 Vertical Operation In Detail 3Ć78 ...
Page 127: ...Appendices Appendices ...
Page 128: ......
Page 170: ...Appendix C Quick Checks Appendices AĆ42 Appendix C Quick Checks Appendices AĆ42 ...
Page 210: ...Appendix F Accessories Appendices AĆ82 Appendix F Accessories Appendices AĆ82 ...
Page 211: ...Glossary Index Glossary Index ...
Page 212: ......
Page 230: ...Index Index IĆ10 Index Index IĆ10 ...
Page 231: ......
Page 232: ......