Specifications
4–6
1710J-Series Waveform Monitors
Table 4–10: Environmental Characteristics
Name
Description
Operating Temperature
0
_
to 50
_
C (+32
_
to 122
_
F)
Storage Temperature
–40
_
to 75
_
C (-40
_
to 158
_
F).
Operating Altitude
To 15,000 feet (4572 meters)
Storage Altitude
To 50,000 feet (15,240 meters)
Vibration
5 minutes at 5 - 15 Hz with 0.060 in. displacement
5 minutes at 15 - 25 Hz with 0.040 in. displacement
5 minutes at 25 - 55 Hz with 0.020 in. displacement
Military Specification: Mil-T-28800D, Paragraph 1.2.2, Class 3
Mechanical Shock
Non-operating: 50 g 1/2 sine, 11 ms duration three shocks per surface (18 total)
Transportation
Qualified under NSTA Test Procedure 1A, Category II (24 inch drop)
Humidity
Will operate at 95% relative humidity for up to five days
Do not operate with visible moisture on the circuit boards
Table 4–11: Certifications and Compliances
EC Declaration of Conformity
1
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 50081-1 Emissions:
EN 55022
Class B Radiated and Conducted Emissions
EN 50082-1 Immunity:
IEC 801-2
Electrostatic Discharge Immunity
IEC 801-3
RF Electromagnetic Field Immunity
IEC 801-4
Electrical Fast Transient/Burst Immunity
Low Voltage Directive 73/23/EEC:
EN 61010-1
Safety requirements for electrical equipment for measurement,
control, and laboratory use
EN 61010-2-031:1994
Particular requirements for hand-held probe assemblies for
electrical measurement and test
EMI
1
FCC 47 CFR Part 15, Sub-part B, Class A
Installation (Overvoltage) Category
Terminals on this product may have different installation (overvoltage) category designations.
The installation categories are:
CAT III
Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.
CAT II
Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected.
CAT I
Secondary (signal level) or battery operated circuits of electronic equipment.
Summary of Contents for 1710J Series
Page 4: ......
Page 14: ...Table of Contents x 1710J Series Waveform Monitors ...
Page 18: ...Service Safety Summary xiv 1710J Series Waveform Monitors ...
Page 22: ...Preface xviii 1710J Series Waveform Monitors ...
Page 23: ...Getting Started ...
Page 24: ......
Page 45: ...Operating Basics ...
Page 46: ......
Page 55: ...Reference ...
Page 56: ......
Page 64: ...Reference 3 8 1710J Series Waveform Monitors ...
Page 65: ...Specifications ...
Page 66: ......
Page 74: ...Specifications 4 8 1710J Series Waveform Monitors ...
Page 76: ......
Page 77: ...Theory of Operation ...
Page 78: ......
Page 100: ...Theory of Operation 5 22 1710J Series Waveform Monitors ...
Page 101: ...Performance Verification ...
Page 102: ......
Page 120: ...Performance Verification 6 18 1710J Series Waveform Monitors ...
Page 121: ...Adjustment Procedure ...
Page 122: ......
Page 138: ...Adjustment Procedure 7 16 1710J Series Waveform Monitors ...
Page 139: ...Maintenance ...
Page 140: ......
Page 169: ...Replaceable Electrical Parts ...
Page 170: ......
Page 204: ...Replaceable Electrical Parts 9 34 1710J Series Waveform Monitors ...
Page 205: ...Diagrams ...
Page 206: ......
Page 226: ......
Page 227: ...Replaceable Mechanical Parts ...
Page 228: ......
Page 240: ...Replaceable Mechanical Parts 11 12 1710J Series Waveform Monitors ...
Page 244: ...Glossary Glossary 4 1710J Series Waveform Monitors ...
Page 251: ......
Page 252: ......