Performance Verification
1710J-Series Waveform Monitors
6–7
5. Check that no falling transition between the 10
m
s and the 110
m
s graticule
marks is more than 1% (0.5 minor division) from a major graticule mark.
6. Select 1 LINE sweep (push and hold).
7. Use the HORIZONTAL position control to place the first calibrator
transition on the 5
m
s graticule mark (left side graticule timing mark).
8. Check for five full cycles of calibrator signal in the center ten major graticule
divisions,
±
2% (one minor division).
9. Select 2 LINE sweep and turn on the HORIZONTAL MAG.
10. Check for one full cycle of calibrator signal in the center ten divisions of the
graticule,
±
2% (one minor division).
11. Use a 75
W
coaxial cable to connect the television signal generator multi-
burst output to the 1710J-Series Waveform Monitor CH A input.
12. Use a 75
W
coaxial cable to connect the other side of the loop-through to the
digital counter input.
13. Adjust the multiburst Frequency for a 5 MHz sine wave as measured on the
digital counter.
14. Select the 1710J-Series Waveform Monitor INT REF and 1 LINE sweep.
Turn on the MAG
15. Set the multiburst generator to Composite.
16. Use the 1710J-Series Waveform Monitor HORIZONTAL position control to
display all but the first and last 10% of the sweep.
17. Check for ten cycles over ten graticule divisions,
±
1.5 minor divisions.
18. Switch the multiburst generator to Continuous and check that its frequency,
as measured on the digital counter, remains set at 5 MHz. Return to
Composite.
Summary of Contents for 1710J Series
Page 4: ......
Page 14: ...Table of Contents x 1710J Series Waveform Monitors ...
Page 18: ...Service Safety Summary xiv 1710J Series Waveform Monitors ...
Page 22: ...Preface xviii 1710J Series Waveform Monitors ...
Page 23: ...Getting Started ...
Page 24: ......
Page 45: ...Operating Basics ...
Page 46: ......
Page 55: ...Reference ...
Page 56: ......
Page 64: ...Reference 3 8 1710J Series Waveform Monitors ...
Page 65: ...Specifications ...
Page 66: ......
Page 74: ...Specifications 4 8 1710J Series Waveform Monitors ...
Page 76: ......
Page 77: ...Theory of Operation ...
Page 78: ......
Page 100: ...Theory of Operation 5 22 1710J Series Waveform Monitors ...
Page 101: ...Performance Verification ...
Page 102: ......
Page 120: ...Performance Verification 6 18 1710J Series Waveform Monitors ...
Page 121: ...Adjustment Procedure ...
Page 122: ......
Page 138: ...Adjustment Procedure 7 16 1710J Series Waveform Monitors ...
Page 139: ...Maintenance ...
Page 140: ......
Page 169: ...Replaceable Electrical Parts ...
Page 170: ......
Page 204: ...Replaceable Electrical Parts 9 34 1710J Series Waveform Monitors ...
Page 205: ...Diagrams ...
Page 206: ......
Page 226: ......
Page 227: ...Replaceable Mechanical Parts ...
Page 228: ......
Page 240: ...Replaceable Mechanical Parts 11 12 1710J Series Waveform Monitors ...
Page 244: ...Glossary Glossary 4 1710J Series Waveform Monitors ...
Page 251: ......
Page 252: ......