Connect Stations

Up to six stations may be connected to each room controller using the header on the back of the room
controller. The leads on the header are paired together to form a point on the room controller. This
means that each header has six points to connect stations to, and one point for data bus connections.

With the wire leads on the connector pointing down, point 1 is on the left. The illustration below shows
the six available points on the header. The point connections are ar-, +-, etc., with each group
of +- connections making up one point. Refer to the figures below for default connection points.

Note:

Room controllers sold prior to warranty number N6861530 (April 2017) were sup-

plied with Red and Red/White wires for Point 2 and Blue and Blue/White wires for Point
3 of their station connection header.

Point 1

Point 2

Point 3

Point 4

Point 5

Point 6

Point 7

+

-

+

-

+

-

+

-

+

-

+

-

+

-

Violet

Violet

White

Stripe

Blue

Blue

White

Stripe

Red

Red

White

Stripe

Orange Orange

White

Stripe

Brown Brown

White

Stripe

Green Green

White

Stripe

White Black

Figure 5 - Single Patient Station Default Room (Address 0-29)

FigFigure ure
9: SF123 Ay
Diagram

Copyright

©

TekTone Sound and Signal Mfg., Inc. All Rights Reserved

IL992 Tek-CARE120 Installation Manual

|

15

Installation

Summary of Contents for Tek-CARE120

Page 1: ...Free 800 327 8466 Sales Option 2 Tech Support Option 3 324 Industrial Park Rd Franklin NC 28734 tektone tektone com Fax 828 524 9968 IL992 Rev 9 8 2020 Tek CARE120 Tone Visual Nurse Call Signaling Sy...

Page 2: ...Tone TekTone Sound Signal Mfg Inc assumes no responsibility or liability for any errors or inaccuracies that may appear in this documentation TekTone the TekTone logo Tek ALERT Tek BRIDGE Tek CALL Tek...

Page 3: ...ter Stations 19 Connect Data Bus Wiring 20 Connect Master Bus Wiring 20 Networking Multiple NC120 CE Modules 20 Configuration Setup 21 Connect the LS450 Config Tool Programming Software 21 Using the C...

Page 4: ...ord stations 1 4 jack call cord sta tions duty stations dome lights and more In addition to these items the Tek CARE120 system can be easily integrated with exist ing TekTone nurse call systems nurse...

Page 5: ...acility The monitor can also modify its own annunciation settings and display system faults The NC403TS uses a 22 inch touchscreen and may be mounted using the CL234 wall mount bracket Refer to IL1030...

Page 6: ...ert staff to pending Tek CARE120 system calls The duty station is installed in areas where call annunciation is needed but the full functionality of a master station is not Four LEDs communicate the t...

Page 7: ...each room in the facility and using the points on that room controller to connect to stations installed in that room In this installation strategy stations are normally wired to their default points o...

Page 8: ...CE Module and ensure reliable system per formance the Tek CARE120 Power Calculator will calculate the master load for the module The cal culate may be downloaded at http www tektone com pdf_files gen2...

Page 9: ...Figure 1 NC120 CE Module Connections Copyright TekTone Sound and Signal Mfg Inc All Rights Reserved IL992 Tek CARE120 Installation Manual 9 Installation...

Page 10: ...0 CE port descriptions On the front of the NC120 CE Module you will find a bank of DIP switches These DIP switches are used to configure various options on the CE module and to provide an address for...

Page 11: ...Bus The Tek CARE120 data bus is a two wire serial communications bus There are four ports on the front of the module each capable of driving up to 1 000 feet of data bus and up to 16 LI122 Series room...

Page 12: ...to room controller and connect the room controllers to the data bus in parallel using T taps Connect the positive wire in the data bus cable to the solid white wire pin 13 on the back of the room cont...

Page 13: ...ns may be installed on each port for a total of six master stations per NC120 CE Module To install multiple masters on one master bus trunk line use CT400P5 splitters Maximum tap length off of the mai...

Page 14: ...t 0A and port 0B must be unique but port 0A and port 1A may have addresses duplicated between them Keep detailed notes of the physical addresses and locations of room controllers installed in the faci...

Page 15: ...h each group of connections making up one point Refer to the figures below for default connection points Note Room controllers sold prior to warranty number N6861530 April 2017 were sup plied with Red...

Page 16: ...connected to each point Do not splice multiple stations together on the same point wiring One to One Installations In most installations one room controller will be used as a connection point for all...

Page 17: ...possible connect the patient station to Point 1 Purple and Purple White wires This is the default connection point for patient stations in a single or dual bed installation SF123 Two Button Pull Cord...

Page 18: ...m the SF125 LED panel to the positive terminal of the desired point on the room controller and connect the black wire to the negative terminal of the desired point on the room con troller Make sure th...

Page 19: ...iod opening the switch at any point during the check in period will satisfy the check in requirement For more information about check in see Check In on page 30 Connect NC415AV Master Stations The two...

Page 20: ...tomatically assigned names Once the sys tem is powered on for the first time go to each master station on the system and touch the Menu button on the home screen then touch the This Master button that...

Page 21: ...are the points on the LI122 Series and PM120 Room Controllers The points are the six physical connections on the rear of each controller shown as points 1 6 in the configuration In addition to the ph...

Page 22: ...tions and or SF122 Dual Patient Stations per room or in an area where two or more Code Blue or Emergency switches are connected to the same room controller This scenario will be referred to as a Custo...

Page 23: ...Figure 13 Config Tool Home Page Copyright TekTone Sound and Signal Mfg Inc All Rights Reserved IL992 Tek CARE120 Installation Manual 23 Configuration Setup...

Page 24: ...enu Remove Private Data allows all private patient data such as medications con ditions etc to be stripped from the configuration so that it can be sent to TekTone for troubleshooting help without rev...

Page 25: ...used to interact with the Tek CARE software These buttons appear at the top of editing panes in the Config Tool Edit Before information in an editing pane may be changed click this button to unlock th...

Page 26: ...criptive file name and save the configuration Enabling and Naming Modules Navigate to the Modules page In the Modules list verify that all of the installed NC120 CE Modules installed on your system ar...

Page 27: ...tation name For example a room controller addressed as 12 connected to port 0A or 0B of CE 01 would be assigned a default station name of 0112 If the same room controller address 12 were connected to...

Page 28: ...edit multiple window is open select the Station IDs to be edited Once stations are selected enable the points that have stations connected and choose the correct station type and behavior Mark any unu...

Page 29: ...lls from the stations in the selec ted zone Click Apply to save your changes Repeat for all zones required by the facility Room Level Reset SF123s can be programmed as Reset Buttons to enable the room...

Page 30: ...which the system expects to see a check in input triggered If the check in input is not triggered during the scheduled time period an Inactivity call annunciates on the system Create several check in...

Page 31: ...he Config Tool In the list of masters that appears select the master to be renamed Click the Edit button to unlock the Details pane for editing In the details pane give each master a unique descriptiv...

Page 32: ...s are connected to Refer to the completed programming worksheet that was created during system installation to ensure cor rect programming The most important concept to remember during system configur...

Page 33: ...ns Move Points to Stations Occasionally it may be necessary to assign a point to a different station than it is defaulted to To move points between Station IDs select the point to be moved from the po...

Page 34: ...avior to a Station ID click Stations in the page selection pane Select the Station ID that the new hardware behavior is to be assigned to and click the Edit button in the Details pane Select the hardw...

Page 35: ...editing the dome light annunciation for the Emergency call type Note that the settings shown above would result in a medium flash rate on the red LED of LI122s zoned to display the Emergency call howe...

Page 36: ...the master sta tion If this option is available on your system the gray buttons on the right side of the NC415AV hous ing are used The buttons are labeled accordingly Using the NC415AV Master Station...

Page 37: ...en displays information about the alarming station the type of call the resident name and more Staff requests Stat Attend and Status can also be placed from the details screen by touching the associat...

Page 38: ...sired station will appear To enable or disable check in for a station dial the station touch the Check in button on the bottom of the screen Select any available check in schedules to apply to the sta...

Page 39: ...ation including the Device ID of the current NC415AV the mod ule and port it is connected to the DIP switch address of the NC415AV and the Tek CARE software version Touch the Clean button to temporari...

Page 40: ...Using the Tek CARE120 System Tek CARE120 Programming Worksheets 40 IL992 Tek CARE120 Installation Manual Copyright TekTone Sound and Signal Mfg Inc All Rights Reserved...

Reviews: