Section V
–
Maintenance and Servicing
1830A Calibration Initiate
To initiate a calibration establish a LAN interface web session as described in Section IV.
Select System Calibration from the LAN interface web menu (Refer to Figure 5.6). Select
Perform Calibration.
Figure 5.6 - System Calibration Web Page
1830A Calibration Procedure
Once a calibration is started it must be completed. The automated process will step the
user though an entire calibration of both channels. Unlike Diagnostics the user will not be
able to select specific procedures. The entire process takes less than an hour. Refer to
Figure 5.7 for the first step in the calibration. Detailed procedures and diagrams will be
provided for each step of the calibration (Refer to Figures 5.4 & 5.5).