10 TEGAM WAY • GENEVA, OHIO 44041
440-466-
6100 • FAX 440
-466-
6-1
Section VI
–
Reprogramming
Reprogramming the 1830A
NOTE: Refer to Section IV to establish a connection via the LAN interface.
1.
Establish TCP/IP connect with 1830A (You will need LAN interface).
2.
Download the 1830A_v1.1.18a.bin and 1830A_WebFix.bin files to a folder on your
computer.
3.
Place the Unit in Standby Mode
4.
Open a web browser, and go to the units home page by going to http://{unit IP address}.
5.
Click 'Login' on the left navigation pane.
6.
Enter 'admin' as the login password. This is the default password when the unit ships. It
may be changed for installations.
7.
Click 'Firmware Upgrade' on the left navigation pane.
8.
Using the web interface, upload the `1830A_WebFix.bin' file. This file fixes an issue with
the web interface where files greater than 2Meg in size cannot be uploaded. The upgrade
stops the 1830A webserver for a few seconds, so give the upgrade a little time to happen.
Please note the web browser may show a time out or connection error. This is normal as
the unit needs to restart its internal web server to apply the fixes.
9.
Navigate back to the unit's home page by going to http://{unit IP address}. The unit will
show the web interface once the web server has restarted. You can tell if the upgrade
was applied successfully by the Copyright notice at the bottom of the page, it should show
'V1.1.18' if the fix was installed properly.
10.
Access the 'Firmware Upgrade' interface again, upload the `1830A_v1.1.18a.bin' file. This
file will take a little time to upload, please don't double click on upload button to make it
go faster. The 1830A Application will shut down, and status information will be displayed
on the LCD screen. Allow the update to proceed, the unit will reboot when it is finished.
Once the unit has rebooted, it is ready for use.