Section IV
–
Operating Instructions
NOTE:
The reference resistor is either 100 Ω or 200 Ω. Additionally, in some modes, the
reference resistance voltage is multiplied by a variable to produce a variable resistance
feature. At resistance other the 100 Ω or 200 Ω, the multiplication adds some uncertainty.
Comp Resistor (Read/Write)
Located SETUP
—
Instrument
—
Comp Resistor menu. The compensation resistor R
Comp
is
used to monitor a constant current source that heats the RF compensation sense
thermistors in the sensor. The compensation resistor can be set to 100 Ω or 200 Ω.
Read Avgs (Read/Write)
Located SETUP
—
Instrument
—
Read Avgs menu. This control is defaulted to 1 average,
which is equal to 8 ADC averages. For each displayed measurement the ADC is averaged
8 times. Control is available in firmware version 1.1.28a and greater.
Display (Read/Write)
Located SETUP
—
Instrument
—
Display menu. This control sets the LCD brightness and LCD
contrast.
System Temperature (Read Only)
Located SETUP
—
Instrument
—
System Temperature menu. This indicates the system
temperature of the 1830A.
Serial Number (Read Only)
Located SETUP
—
Instrument
—
Serial Number menu. This indicates the serial number of the
1830A.
Firmware Version (Read Only)
Located SETUP
—
Instrument
—
Firmware Version menu. This indicates the current firmware
version and firmware date on the 1830A.
Analog Board Firmware Version (Read Only)
Located SETUP
—
Instrument
—
Analog Brd FW Ver menu. This indicates the current
firmware version and firmware date for the analog board on the 1830A.
Auto Scale Reading (Read/Write)
Located SETUP
—
Instrument
—
Auto Scale Reading menu. This control is defaulted to False,
all power reading will be displayed in mW. Setting this control to true will show power
reading in scaled notation from nW to kW. Control is available in firmware version 1.1.28a
and greater.
Cal Factor Limit (Read/Write)
Located SETUP
—
Instrument
—
Cal Factor Limit menu. This control limits the maximum cal
factor. By default set top 2.0 can be increased beyond 2.0 in firmware version 1.1.28a
and greater.