PROVASET 2P code 2P2B00 - 2P6B00 - Instruction manual
rev. 14/02/2013
page 23
6.4
"CONFIGURATION AND OPTIONS" menu
Parameter
LANGUAGE
ITALIANO - ENGLISH - DEUTSCH - FRANCAIS - ESPANOL
Parameter
TEST TABLE
Number of the table used for the tests
Range 1
÷
16.
Parameters
TEST PRESSURE
MAX + % and MIN - %
Percentage tolerance limits, %Pr max and %Pr min, in comparison to the Pr pressure of the
operative test table
Range 0
÷
99.
Parameter
PRESSURE SCALE
Unit of measurement relating to pressure.
The programmable parameters "test pressure" (Pr) and "leak limit" (Dp) inside the parameter
tables are automatically converted to the new unit of measurement.
mmH2O - mbar - kPa - mmHg - psi inH2O
The changes do not apply to the test in progress.
Parameter
TEST PHASES
Testing cycle phases
T1 – T2 – T3
: Three- phases leak test.
Pr/T1– T2 – T3
: Three- phases leak test, but with filling phase terminated when the test
pressure is reached or at the end of time T1.
T1 – T2
:
Two- phases leak test.
The changes do not apply to the test in progress.
Parameters
TIME BASE
Counting rate of the T1, T2 and T3 phase times.
0:
0.1s
1:
1s
Parameter
START COMMAND
The device can accept a "START" command from the Start key on the front panel and from
the external Start contact in connector "D", picture 2 point 7, of the rear panel.
0:
Manual, from the Start key or the external Start contact in parallel.
1:
External, only from the external Start contact, not from the Start key.
CONFIGURATION ->
Language:
Test table:
Test pressure:
[MAX + %] XXX
[MIN - %] YYY
CONFIGURATION ->
Pressure Scale:
Test Phases:
Time base:
CONFIGURATION ->
Start Command:
Enable Test for
Obstruction:
Final Pressure
Discharge:
[
↓
]
[
↓
]
CONFIGURATION
Buzzer:
Programming:
[table n. ] XX
[parameters] XX
[
↓
]