PROVASET 2P code 2P2B00 - 2P6B00 - Instruction manual
rev. 14/02/2013
page 14
4.1.2
Testing cycle procedure
The tester accepts test starting commands at any moment and during the visualisation of
any menu page, apart from those relating to the pressure sensor calibration.
The test table used when testing is directly displayed when starting or after a "reset" of the
test in progress.
At the end of a test or during the tests, the test table can be displayed in the message menu
(see par. 5.1) and can be selected (figure from 1
÷
16) in the "Test table selection" menu (see
par. 6.2).
Any change of the test tables carried out during a test in progress will be active only for the
next test.
4.1.3
Three-phases testing cycle (T1, T2, T3)
Ready and waiting for the next test, that will be carried out
with the displayed table no. XX.
The green or the red light is blinking, depending on the result
of the last test performed.
Only for
2P2B00
model
In case the manual pressure discharge option was selected,
the green and the red lights are both off.
Test running:
Filling time T1
xx.x
: Remaining T1 time (decreasing value).
yyyy
: Line pressure read by the pressure sensor.
The valves, opened when starting, are closed at the end of
time T1.
Test running:
Settling time T2
xx.x
: Remaining T2 time (decreasing value).
yyyy
: Line pressure read by the pressure sensor.
Test running:
Time T3, measurement of the pressure decay Dp
xx.x
: Remaining T3 time (decreasing value).
yyyy
: Pressure decay detected (with sign).
TEST
TABLE n XX
TEST: FILLING
T1 s XX.X
mmH2O YYYY
TEST
TABLE n XX
TEST: Dp MEASURE
T3 s XX.X
mmH2O - YYYY
READY
TABLE n XX
T1 xx.x mmH20
T2 xx.x Pr
YYYY
T3 xx.x Dp
XX
TEST
TABLE n XX
TEST: SETTLING
T2 s XX.X
mmH2O YYYY