PROVASET 2P code 2P2B00 - 2P6B00 - Instruction manual
rev. 14/02/2013
page 22
6.2
"TEST TABLE SELECTION" select menu
Select the number of the operative test table for the tests with the
[+]
/
[-]
keys, by confirming
with key
[P
↵
]
.
xx
: NUMBER OF THE OPERATIVE PARAMETER TABLE
Range 1
÷
16
NOTE:
The number of the operative test table can be programmed also inside the
"CONFIGURATION AND OPTIONS" menu (see par. 6.1)
6.3
"TEST TABLES" modify menu
Select the number of the test table to be changed.
Select the parameter to be changed with the
[
↓
]
key.
Change the numerical selected parameter with the
[+]
key to increase or with the
[-]
key to
decrease and confirm the change (blinking parameter) with button
[P
↵
]
.
PARAMETERS
DESCRIPTION
RANGE
T1
FILLING PHASE
0.0
÷
99.9 s or 0
÷
999 s
T2
SETTLING PHASE
0.1
÷
99.9 s or 1
÷
999 s
T3
MEASUREMENT PHASE
0.1
÷
99.9 s or 1
÷
999 s
Pr
TEST PRESSURE
SCALE RANGE
Dp
PRESSURE DELTA
SCALE RANGE
PROVASET
2P
– Test parameters table and programmable ranges
SCALE
2P2B00
RANGE
2P6B00
RANGE
mmH2O 0
÷
20394 0
÷
61182
mbar
0
÷
2000.0 0
÷
6000.0
kPa
0
÷
200.00 0
÷
600.00
mmHg 0
÷
1500.0 0
÷
4500.0
PSI
0
÷
30.00
0
÷
90.00
inH2O
0
÷
802,5 0
÷
2407,5
PROVASET
2P
–Programmable pressure scales and ranges
SELECT
TABLE n. XX
T1 xx.x mmH20
T2 xx.x Pr
YYYY
T3 xx.x Dp
XX
PROGRAMMING
TEST TABLE
SELECTION
[
↓
]
PROGRAMMING
TEST TABLES
[
↓
]
MODIFY
TABLE n. XX
T1 xx.x mmH20
T2 xx.x Pr
YYYY
T3 xx.x Dp
XX