PROVASET 2P code 2P2B00 - 2P6B00 - Instruction manual
rev. 14/02/2013
page 16
4.4
Test results
The test is ended with a
failed
result if, at the end of the "Dp
measure" time, the measured pressure decay - as absolute
value without sign - is equal or higher than the set Dp
maximum limit.
The red led blinks up to the beginning of a new test.
The test is ended with a
good
result if, at the end of the "Dp
measure" time, the measured pressure decay - as absolute
value without sign - is lower than the set Dp maximum limit.
The green led blinks up to the beginning of a new test.
If the manual pressure discharge option was selected, the
pressure drops down when the operator disconnects the
tested part, and then the green light turns off and the display
changes to the message indicating ready for a new test with
the selected table.
4.5
Alarm indications:
Causes and solutions
Under alarm conditions, the test in progress is cancelled.
- TABLE No. XX
Error in the parameters stored in the eeprom.
Reprogram the test table.
- CONFIGURATION
Error in the parameters stored in the eeprom.
Reprogram
all
the parameters of the "CONFIGURATION AND OPTIONS" menu.
- CALIBRATION
Error in the calibration constants of the pressure transducer.
Carry out the calibration procedure.
- PRESSURE
Error in the reading of the pressure transducer.
It is necessary to repair the device.
NOTE
:
Until these operations are not carried out in the correct way, the device won’t
be ready to start the test.
END OF TEST
TABLE n XX
result:
FAILED
mmH2O Dp
XXXXX xxx
END OF TEST
TABLE n XX
result:
PASSED
mmH2O Dp
XXXXX xxx
ALARMS
- TABLE n XX
- CONFIGURATION
- CALIBRATION
- PRESSURE