8 - 2
Te-Stack Operating Manual, 391862, en, V2.3
8 - Troubleshooting
Troubleshooting Table
8.1 Troubleshooting Table
Trouble-
shooting by the
Operator
The following table lists problems and errors and instructs how to eliminate them.
Tab. 8-1
Troubleshooting table
Problem / error
Possible cause
Corrective measures
Te-Stack (base unit)
Microplates dropped or
picked up incorrectly.
The Te-Stack position
is incorrect.
The microplates do
not meet the SPS
standard.
The plate type specifi-
cation is incorrect.
Teach the position.
Refer to the “Instrument Software Manual”
Always use microplates meeting the SPS
standard.
Correct the specification.
Stack for microplates
Microplate gets stuck
Microplates are
loaded incorrectly.
The grippers got
stuck.
The solenoid is defec-
tive.
Adjust the microplates.
Check the gripper mechanism.
If the problem cannot be solved, contact the
Tecan FSE.
Contact the Tecan FSE.
Stack for DiTi boxes
DiTi box gets stuck.
DiTi boxes are loaded
incorrectly.
The grippers got
stuck.
The solenoid is defec-
tive.
Adjust the DiTi boxes.
Check the gripper mechanism.
If the problem cannot be solved, contact the
Tecan FSE.
Contact the Tecan FSE.
Summary of Contents for Te-Stack 10612251
Page 1: ...Te Stack Operating Manual 391862 en V2 3 Operating Manual Te Stack DiTi Box and Plate Stacker ...
Page 6: ...0 IV Te Stack Operating Manual 391862 en V2 3 0 Preface FCC Rules ...
Page 10: ...iv Te Stack Operating Manual 391862 en V2 3 Table of Contents ...
Page 14: ...1 4 Te Stack Operating Manual 391862 en V2 3 1 About This Manual Abbreviations ...
Page 24: ...2 10 Te Stack Operating Manual 391862 en V2 3 2 Safety General Safety Rules ...
Page 62: ...6 10 Te Stack Operating Manual 391862 en V2 3 6 Operation Defining Scripts ...
Page 74: ...9 4 Te Stack Operating Manual 391862 en V2 3 9 Shutdown Storage and Shipping Transport ...
Page 82: ...11 6 Te Stack Operating Manual 391862 en V2 3 11 Spare Parts and Accessories Options ...
Page 86: ...12 4 Te Stack Operating Manual 391862 en V2 3 12 Customer Support ...