6 - 6
Te-Stack Operating Manual, 391862, en, V2.3
6 - Operation
Tasks During a Run
ATTENTION
Risk of damage to equipment. Removing stacker column loaded with nested DiTi
will lead to damage to DiTi.
Install the carrier slide before removing the column.
Or
Remove all nested DiTi stacks form the columns before removal.
Fill the microplates or the DiTi racks from the top into the stacker column.
6.5.3 Refilling Stacker Column for DiTi Boxes
ATTENTION
Compatibility.
A stacker column for DiTi boxes cannot be used for microplates.
DiTi boxes always have to be used without lid.
Fill the DiTi boxes from the top into the stacker column.
6.5.4 Emptying the DiTi Waste (Te-MO Version Only)
1
Click the
Pause
button.
2
Replace the full waste bag by an empty one.
3
Fix the waste bag.
4
Click the
Continue
button.
Fig. 6-3
Overview of Freedom EVO version in front position
Summary of Contents for Te-Stack 10612251
Page 1: ...Te Stack Operating Manual 391862 en V2 3 Operating Manual Te Stack DiTi Box and Plate Stacker ...
Page 6: ...0 IV Te Stack Operating Manual 391862 en V2 3 0 Preface FCC Rules ...
Page 10: ...iv Te Stack Operating Manual 391862 en V2 3 Table of Contents ...
Page 14: ...1 4 Te Stack Operating Manual 391862 en V2 3 1 About This Manual Abbreviations ...
Page 24: ...2 10 Te Stack Operating Manual 391862 en V2 3 2 Safety General Safety Rules ...
Page 62: ...6 10 Te Stack Operating Manual 391862 en V2 3 6 Operation Defining Scripts ...
Page 74: ...9 4 Te Stack Operating Manual 391862 en V2 3 9 Shutdown Storage and Shipping Transport ...
Page 82: ...11 6 Te Stack Operating Manual 391862 en V2 3 11 Spare Parts and Accessories Options ...
Page 86: ...12 4 Te Stack Operating Manual 391862 en V2 3 12 Customer Support ...