41
Total Control Systems
www.tcsmeters.com
ICON
INDICATION
SOLUTION
Printer is connected with paper
Printer is Enabled, connected and with paper. Awaiting delivery.
Printer is out of paper
Printer Enabled, connected, but without paper.
No printer connection
Printer has no communication with TCS Register
Check cable for damage
Epson printer DIP switch should be on 3
Power supply fuse is blown
Valve or Pump not open in
Delivery
Valve or Pump are closed within delivery. Pump Starter delay timing
can be adjusted under W&M Settings.
Active delivery
Valve or Pump Enabled
Multiple Tank Delivery
Multiple Tank Delivery is Enabled and begun by pressing Shift + Start
You’re in System Settings
Screen
Exit out and go to the Delivery Screen
You’re in Weights & Measure
Settings
Only shows when in W&M Settings. To remove, make sure the cali-
bration screw on the register is screwed in.
AIR
Air is present in system or level
sensor is low or high
The Air Eliminator Detection is Enabled.
Density is out of range
The Density reading is outside the limit entered values
Water is out of range
Water PPM reading is outside the limit entered values
BASIC DISPLAY ICON INDICATION
We will categorize Troubleshooting with HARDWARE/SOFTWARE, SYSTEM, WEIGHTS & MEASURES and
PRODUCT Settings in the TCS 3000 Register. TCS HUB and Modem Settings follow register troubleshooting.
The list is not all
-
inclusive and should only be used as a guide.
PROBLEM
PROBABLE CAUSE
SOLUTION
Unit will not power up
or there is no display.
Inadequate supply voltage,
+9 to 28VDC are required
for operation.
Ground short on terminals,
wire harness or lug.
1.
With the key in the accessory position, check the battery voltage to
the circuit board. Use the negative terminal as DC ground. While
the TCS3000 will power
-
up at +9VDC, it is recommended that the
input be at least +12.6VDC.
2.
There is a ground short within the register or on the cable harness.
Check wiring on grounds in register or chassis. Some ground shorts
come from a cable jacket being worn through with a tie
-
down.
3.
Check the 7.5A, in
-
line fuse for continuity. It is located in the acces-
sory power line. Replace if necessary.
4.
If the green light blinks fast it is doing what it is supposed to, but
then it will go to a slow blink.
If you see red, then you do not have
enough power on the register.
It must see a minimum of 9vdc.
5.
If the red light comes on and the unit doesn’t power
-
up. Check the
voltage if the voltage is OK but the temperature is
-
20 F this will
indicate that the heater is warming the display to prevent any dam-
age to the VGA display and should come
-
up shortly.
HARDWARE/SOFTWARE TROUBLESHOOTING