6-3
TTi
®
Colorimeter User’s Manual
Part 6
Before You Call for Service
ERROR CODES
Error codes are displayed when the instrument detects misuse or failure of software, hardware, or batteries.
When an Error Code message is displayed, use the table below to determine the likely cause and user action.
DO NOT attempt to service the instrument yourself, other than the User Action described in the table.
There are no user-serviceable parts other than the batteries. Any attempt to open the instrument will void the
warranty.
To contact Tech Support, call
800
-837-8548 between 8:00 a.m.–4:45 p.m. (Eastern Time), Monday through
Friday, except major holidays.
Display Message
Likely Cause
User Action
Error: 001
420 nm error.
See
User’s Manual.
Obstruction in the sample cell compartment
during initialization.
420 nm LED or detector error.
Check for a sample cell or foreign object in
the sample cell compartment. Remove; then
restart device.
Press any key to override and allow testing at
other wavelengths.
Contact Tech Support if necessary.
Error: 002
470 nm error.
See
User’s Manual.
Obstruction in the sample cell compartment
during initialization.
470 nm LED or detector error.
Check for a sample cell or foreign object in
the sample cell compartment. Remove; then
restart device.
Press any key to override and allow testing at
other wavelengths.
Contact Tech Support if necessary.
Error: 003
520 nm error.
See
User’s Manual.
Obstruction in the sample cell compartment
during initialization.
520 nm LED or detector error.
Check for a sample cell or foreign object in
the sample cell compartment. Remove; then
restart device.
Press any key to override and allow testing at
other wavelengths.
Contact Tech Support if necessary.
Error: 004
570 nm error.
User’s Manual.
Obstruction in the sample cell compartment
during initialization.
570 nm LED or detector error.
Check for a sample cell or foreign object in
the sample cell compartment. Remove; then
restart device.
Press any key to override and allow testing at
other wavelengths.
Contact Tech Support if necessary.
Error: 005
620 nm error.
See
User’s Manual.
Obstruction in the sample cell compartment
during initialization.
620 nm LED or detector error.
Check for a sample cell or foreign object in
the sample cell compartment. Remove; then
restart device.
Press any key to override and allow testing at
other wavelengths.
Contact Tech Support if necessary.
Error: 006
660 nm error.
See
User’s Manual.
Obstruction in the sample cell compartment
during initialization.
660 nm LED or detector error.
Check for a sample cell or foreign object in
the sample cell compartment. Remove; then
restart device.
Press any key to override and allow testing at
other wavelengths.
Contact Tech Support if necessary.
Summary of Contents for K-8000
Page 2: ......
Page 3: ...TTi Colorimeter Series User s Manual...
Page 4: ......
Page 10: ...1...
Page 11: ...1 1 TTi Colorimeter User s Manual Part 1 General Information 1 Part 1 General Information...
Page 12: ......
Page 19: ...2 1 TTi Colorimeter User s Manual Part 2 Testing Primer 2 Part 2 Testing Primer...
Page 20: ......
Page 27: ...3 1 TTi Colorimeter User s Manual Part 3 Colorimeter Operation 3 Part 3 Colorimeter Operation...
Page 28: ......
Page 75: ...4 1 TTi Colorimeter User s Manual Part 4 Using the PC App 4 Part 4 Using the PC App...
Page 76: ......
Page 81: ...5 1 TTi Colorimeter User s Manual Part 5 Maintenance 5 Part 5 Maintenance...
Page 82: ......
Page 86: ......
Page 90: ......
Page 92: ......
Page 95: ...8 1 TTi Colorimeter User s Manual Part 8 Compliance Synopsis 8 Part 8 Compliance Synopsis...
Page 96: ......
Page 99: ......
Page 100: ...Part 5547...