3-14
TTi
®
Colorimeter User’s Manual
Part 3
Colorimeter Operation
STEP 4
Prepare blank.
A blank is used to ZERO the instrument and establish a reference
point for reading samples. A blank may or may not contain reagents
depending on specific test requirements. Instructions supplied with
each test will describe how to prepare the blank.
Here is a typical procedure for preparing a blank:
• Rinse and fill 25 mm sample cell to 10 mL mark with sample;
then cap.
• Insert sample cell into sample cell compartment. Align marks.
STEP 5
Zero the Colorimeter.
Select ZERO
using
; then press ENTER .
The ZEROING transition graphic will be displayed; then zero (0.00
for this example) will be displayed.
This zero reference point will be stored for all subsequent readings
until the test is exited. The instrument can be re-zeroed at any time
during the test procedure.
Align arrow on
sample cell with
notch on meter.
Summary of Contents for K-8000
Page 2: ......
Page 3: ...TTi Colorimeter Series User s Manual...
Page 4: ......
Page 10: ...1...
Page 11: ...1 1 TTi Colorimeter User s Manual Part 1 General Information 1 Part 1 General Information...
Page 12: ......
Page 19: ...2 1 TTi Colorimeter User s Manual Part 2 Testing Primer 2 Part 2 Testing Primer...
Page 20: ......
Page 27: ...3 1 TTi Colorimeter User s Manual Part 3 Colorimeter Operation 3 Part 3 Colorimeter Operation...
Page 28: ......
Page 75: ...4 1 TTi Colorimeter User s Manual Part 4 Using the PC App 4 Part 4 Using the PC App...
Page 76: ......
Page 81: ...5 1 TTi Colorimeter User s Manual Part 5 Maintenance 5 Part 5 Maintenance...
Page 82: ......
Page 86: ......
Page 90: ......
Page 92: ......
Page 95: ...8 1 TTi Colorimeter User s Manual Part 8 Compliance Synopsis 8 Part 8 Compliance Synopsis...
Page 96: ......
Page 99: ......
Page 100: ...Part 5547...