3-22
TTi
®
Colorimeter User’s Manual
Part 3
Colorimeter Operation
STEP 1
Select AC
using
; then press ENTER .
STEP 2
Adjust the displayed result
(2.90 ppm Cl
2
) to the known
concentration of the calibration standard (3.00 ppm Cl
2
) using
; then press ENTER . The numbers will continuously scroll
when
arrows are held down.
The display will return to the RESULT screen and the displayed
result will agree with the concentration of the calibration standard
(3.00 ppm Cl
2
). The AC icon will be displayed in the title bar and
all test results for this test will now be calculated using the
Adjust Calibration factor until the Adjust Calibration option is
turned off.
Limit Alert:
If the result is adjusted to a value beyond a predeter-
mined limit, either
↑
LIMIT or
↓
LIMIT will be displayed. Reversing
the scroll direction will return the display to a numeric value.
Summary of Contents for K-8000
Page 2: ......
Page 3: ...TTi Colorimeter Series User s Manual...
Page 4: ......
Page 10: ...1...
Page 11: ...1 1 TTi Colorimeter User s Manual Part 1 General Information 1 Part 1 General Information...
Page 12: ......
Page 19: ...2 1 TTi Colorimeter User s Manual Part 2 Testing Primer 2 Part 2 Testing Primer...
Page 20: ......
Page 27: ...3 1 TTi Colorimeter User s Manual Part 3 Colorimeter Operation 3 Part 3 Colorimeter Operation...
Page 28: ......
Page 75: ...4 1 TTi Colorimeter User s Manual Part 4 Using the PC App 4 Part 4 Using the PC App...
Page 76: ......
Page 81: ...5 1 TTi Colorimeter User s Manual Part 5 Maintenance 5 Part 5 Maintenance...
Page 82: ......
Page 86: ......
Page 90: ......
Page 92: ......
Page 95: ...8 1 TTi Colorimeter User s Manual Part 8 Compliance Synopsis 8 Part 8 Compliance Synopsis...
Page 96: ......
Page 99: ......
Page 100: ...Part 5547...