TP350 User’s Manual
●Chapter 2: Installation Requirements
1. Installation Conditions
To ensure optimum performance and long and stable product life, observe the
following precautions when selecting an installation location.
Note that installing the system in a location that does not meet the requirements
listed below can result in malfunctions and/or system breakdown. Replacement of
components and system repairs for any deterioration in performance or damage
caused by failure to adhere the following conditions will not be cover by warranty.
1. Install system in a comparatively dust-free location
2. Avoid locations subject to corrosive or flammable gases.
3. Avoid locations subject to either strong shocks or continuing low levels of
vibration.
4. Install the system away from direct sunlight.
5. Install the system as far away as possible from electrical equipment that
generates strong magnetic fields, electrical fields and high frequencies. Avoid
locations near equipment that generate noise caused by electrical motors and
transformers.
6. Choose a flat location capable of withstanding the design weight of the
system.
7. The installation environment must be within 15 to 30
℃
temperature range and
20 to 80
%
relative humidity range.
8. Avoid locations subject to sharp temperature fluctuations, exhaust from
air-conditioning systems and condensation caused by sharp humidity
fluctuations.
9. Avoid locations air vents located on the right and left and rear sides. (Leave a
space of the both sides at least 15cm and the backside at least 10cm)
10. Choose an operating location with good ventilation.
11. Do not operate the system in a cold room.
12. Select a room that can provide sufficient power (frequency, voltage and
current) for system operation.
13. Do not attempt to move the system while holding it by the cable of backside,
as this can damage to components. Doing so can also result in personal injury
due to part breakage resulting in the system falling down.
14. Do not open the lid for the PCR reactions. It is expected to decrease the
accuracy of temperature and generate the condensation at the low
temperature (below 10
℃
)