AR 2000 Operator's Manual
6 2
Preventing Solvent Evaporation
If you are using samples that contain volatile solvents or are water-based, evaporation can cause problems
during measurements. TA Instruments has overcome this problem by using a solvent trap cover, which sits over
the geometry (but does not touch it).
Solvent trap version geometries have a well on top of the geometry. Place a small amount of the relevant solvent
into this well. The solvent trap cover has a lip that sits in the solvent, allowing the free space around the sample
to become saturated with the solvent vapor, which prevents evaporation.
A schematic of a solvent trap cover and geometry is shown in Figure 6.7 below.
When using solvent trap systems it is generally advisable to run the inertia correction wizard (with solvent in
the trap, but no sample loaded).
Preventing Slippage at
Sample/Geometry Interface
Some samples, such as hydrogels, contain a lot of water that can migrate to the surface of the sample. This can
cause a film layer to form between the bulk of the material and the geometry surface, causing slippage at this
interface. To alleviate this problem, use special crosshatched geometries, which, in effect, have the measuring
surface slightly roughened. (However, when you use these
crosshatched
geometries, there is a trade-off between
absolute accuracy and repeatability.)
Solvent well
Solvent trap
version
geometry
Solvent
vapor
saturated
free space
Figure 6.7
Solvent Trap Cover and Geometry
Summary of Contents for AR 2000
Page 24: ...AR 2000 Operator s Manual 24...
Page 30: ...AR 2000 Operator s Manual 30...
Page 34: ...AR 2000 Operator s Manual 34...
Page 66: ...AR 2000 Operator s Manual 66...
Page 84: ...AR 2000 Operator s Manual 84...
Page 120: ...AR 2000 Operator s Manual 120...
Page 126: ...AR 2000 Operator s Manual 126...
Page 128: ...AR 2000 Operator s Manual 128...
Page 132: ...AR 2000 Operator s Manual 132...
Page 134: ...AR 2000 Operator s Manual 134...