AR 2000 Operator's Manual
7 0
Installing the (Optional) Vortex Air Cooler
Follow the steps below to attach and connect the vortex air cooler to the AR 2000. Refer to the figures as needed.
1.
Use the two screws provided to mount the vortex air
cooler bracket to the rear of the AR 2000 casting as
shown in Figure 7.5 to the right.
2.
Clip the vortex air cooler into the spring clips with the
brass muffler extending upward as shown in Figure 7.6
below.
3. Remove the metal push-fit connector from
the inlet port on the UHP and fit the
Swagelok adapter supplied in the kit. See
Figure 7.7 to the right. (Note that once this
has been fitted, it cannot be removed.
Returning to the push-fit connector will
require the supplied adapter.)
Figure 7.5
Mounting the Bracket
Mounting
Screws
(STEP 1)
Air Inlet
From
2-Way
Valve
Figure 7.6
Attaching the Vortex Air Cooler
Exhaust
Vent
(STEP 6)
Exhaust Air
From UHP
(STEP 5)
Muffler
Vortex Tube
Cold Air
Oultet to
UHP
White
Tubing
(STEPS
7&8)
Black Insulated Tube
(STEP 4)
Figure 7.7
UHP Swagelok Adapter
Insulated
Tube
Swagelok
Adapter
(STEP 3)
(STEP 4)
(STEP 5)
White Tubing
Summary of Contents for AR 2000
Page 24: ...AR 2000 Operator s Manual 24...
Page 30: ...AR 2000 Operator s Manual 30...
Page 34: ...AR 2000 Operator s Manual 34...
Page 66: ...AR 2000 Operator s Manual 66...
Page 84: ...AR 2000 Operator s Manual 84...
Page 120: ...AR 2000 Operator s Manual 120...
Page 126: ...AR 2000 Operator s Manual 126...
Page 128: ...AR 2000 Operator s Manual 128...
Page 132: ...AR 2000 Operator s Manual 132...
Page 134: ...AR 2000 Operator s Manual 134...