AR 2000 Operator's Manual
121
Chapter 10
Do's and Don'ts
Overview
Please read this chapter thoroughly before using your rheometer. It may be helpful to prepare a copy of these
points and place them in a prominent position near the instrument.
DO
...
ensure the air supply is always very clean at a stable pressure. A coarse filter and desiccant dryer system
and double filter regulator assembly should be used.
...
look out for water collection in the filter bowls before and during use of the instrument. Drain the filter
bowls whenever necessary. (See Chapter 5.)
...
look for visible signs of dirt on the filter elements in the filter regulator before use. Replace whenever
necessary.
...
replace the air-bearing clamp when the air in ON, and always ensure it is in place when the instrument is
moved or when the air is switched off.
...
refit the air connector plug supplied with the instrument whenever the air line is disconnected.
...
disconnect and blast air through the air line whenever starting up after any period which air has been
switched off. (Not necessary every morning, unless you know your air supply tends to accumulate water
overnight.)
...
connect and switch on air supply before switching the instrument on.
...
switch on water supply to the Peltier plate before switching the instrument on.
...
use good laboratory practice when using the instrument. Wear safety glasses and protective clothing
where necessary.
Summary of Contents for AR 2000
Page 24: ...AR 2000 Operator s Manual 24...
Page 30: ...AR 2000 Operator s Manual 30...
Page 34: ...AR 2000 Operator s Manual 34...
Page 66: ...AR 2000 Operator s Manual 66...
Page 84: ...AR 2000 Operator s Manual 84...
Page 120: ...AR 2000 Operator s Manual 120...
Page 126: ...AR 2000 Operator s Manual 126...
Page 128: ...AR 2000 Operator s Manual 128...
Page 132: ...AR 2000 Operator s Manual 132...
Page 134: ...AR 2000 Operator s Manual 134...