CAUTION: Although the pressure gauge on the cell may to appear to register zero, the
level of pressure in the cell may still be above ambient.
•
Disconnect the pressure cell from the manifold at the designated disconnection point at bottom of the
angled pipe.
•
Raise the head and decouple the magnet assembly and the rotor, and remove the Peltier jacket from the
rheometer using the procedures outlined previously in "Step 4: Loading a Sample."
•
Remove the cup and rotor for cleaning. See "Maintaining the Cell" for details.