AR 2000 Operator’s Manual
90
Magnet Assembly
The magnet assembly attaches to the rheometer's rotating spindle,
and then lowers over the rotor assembly. The spindle collar of the
magnet assembly includes an insert adapter. The adapter insert
should remain in the collar for use with AR2000 or AR2000ex
rheometers. If using the pressure cell with an AR-G2 rheometer,
take the adapter insert out of the spindle collar by removing the
two Phillips (or cross head) screws extending for the outer surface
of the spindle collar. See the figure below. Like the rotor assembly,
the magnet assembly contains a 4-pole magnet. When the spindle
and magnet assembly are rotated, the attraction between the two
4-pole magnets produces a corresponding rotation of the rotor.
There is no physical contact between the two assemblies.
CAUTION: Do not place magnetic storage
media near the magnet assembly, as it con-
tains a powerful magnet capable of destroying
magnetically recorded material.
Figure 8.6
Magnet Assembly with
Cover
Reference
Mark
Figure 8.7
Magnet Assembly with AR 2000/
AR 2000ex Adapter Insert
Adapter
Insert
(Used with
AR 2000/AR 2000ex.)
(Remove for AR-G2.)
Summary of Contents for AR 2000
Page 24: ...AR 2000 Operator s Manual 24...
Page 30: ...AR 2000 Operator s Manual 30...
Page 34: ...AR 2000 Operator s Manual 34...
Page 66: ...AR 2000 Operator s Manual 66...
Page 84: ...AR 2000 Operator s Manual 84...
Page 120: ...AR 2000 Operator s Manual 120...
Page 126: ...AR 2000 Operator s Manual 126...
Page 128: ...AR 2000 Operator s Manual 128...
Page 132: ...AR 2000 Operator s Manual 132...
Page 134: ...AR 2000 Operator s Manual 134...