deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 96/139
www.systec-controls.de
Calibration
Your deltawaveC-P provides a calibrating function that can be used to
calibrate the flow. It may also be used to calibrate the analog outputs and
PT100 temperature sensors.
How to access the "Calibration" menu:
In the primary measuring window "Flow 1": Select "SETUP" -> "CMPL
SETUP" -> Navigate to "Calibration Setup" in the main menu
“
11.4. Calibrating flow
Select "Flow" from the calibration menu“.