deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 86/139
www.systec-controls.de
11.
Parameterize I/O
How to access the I/O setup menu
In the primary measuring window "Flow 1
":
Select "SETUP" -> "CMPL
SETUP". In the „complete setup“ menu, select "I/O Setup“
11.1.
Parameterizing the 4 mA to 20 mA current outputs
Your deltawaveC-P provides two 4 mA to 20 mA current outputs. These
outputs can be assigned different measured values. The outputs are set
actively by default. This means that your deltawaveC-P always provides a
voltage at these outputs.
Any connection of an active 4 mA to 20 mA analog output
of deltawaveC-P with an external device that also provides
a voltage at its inputs will cause fatal damage to your
deltawaveC-P and external device. Before you
interconnect both devices,
always verify that your
external recording (PCS) system is set to passive
state!