deltawaveC-P
User manual
systec Controls Mess- und Regeltechnik GmbH 31/139
www.systec-controls.de
7.2.
Mounting positions for transducers
7.2.1. Basics on the mounting of ultrasonic transducers
The pipe always has to be filled completely at the
mounting positions of the ultrasonic transducers! It is
not possible to take measurements on partially filled
piping
The ultrasonic transducer can be operated in any mounting position.
However, conformation with the mounting positions shown below is
mandatory: The drawing shows the side view of the piping
7.2.2. Mounting ultrasonic transducers on horizontal piping
Picture 10 Preferable mounting positions for ultrasonic transducers (1)
On horizontal piping, it is recommended to mount the transducer with an
offset of approx. +/-45% to the horizontal plane. This is based on the fact
that there is a risk of the accumulation of bubbles in the upper section and
sedimentation in the lower section of the pipe.