27
Remove the insertion instruments.
Using the hexagonal screwdriver
or ratchet wrench with socket
extension, align the end cap
with the nail axis to prevent
cross-threading. Thread the
green end cap into the nail
until it seats fully.
As the final threads of the end
cap turn into the nail, the
surgeon will feel increased
resistance. Continue turning
until the shoulder of the end
cap contacts the proximal
nail end. This will help
prevent backout.
Note:
For 13 mm –15 mm nails
with Standard Locking Sleeve,
use a green 15 mm Titanium
End Cap [474.92x].
Use the ratchet wrench to insert the
green end cap
End cap fully seated
ST
ANDARD
L
OCKING
Use the hexagonal screwdriver to
insert the green end cap