Enlarge the
femoral opening
Fully insert the broach into
the femoral opening using
a twisting hand motion, to
accommodate the 15 mm
diameter 130° locking sleeve.
1
Assemble the implants
Orient the pink 130° locking
sleeve to the nail so the word
“lateral” faces laterally (convex
side of nail bow faces anteriorly).
Seat the sleeve onto the proximal
nail end.
2
31
LATERAL
130°
130
˚
ANTEGRADE
LOCKING
Note:
For 13 mm –15 mm nails,
use 130° Locking Sleeve
[474.913].
130
°
Antegrade Locking