45
44
Remove the trial cup and trial humeral implant using the
humeral implant driver.
select the appropriate final modular humeral implants
that correspond to the trial implants. place the final
modular epiphysis on the final modular stem in the
same rotational position used for the trial implants
(Figure 80).
screw the final modular epiphysis together with the
final humeral stem, using the 3.5 mm hex screwdriver
(yellow handle) and the special locking wrench for
modular implants (size 10-12 or 14-16) (Figure 81).
Both components should then be mounted on
the humeral implant driver and driven down the
intramedullary canal, aligning the lateral fin of the stem
with the bicipital groove (Figure 82). The epiphysis border
should be aligned with the bone resection border.
DEFiNiTivE humERAL impLANTs iNsERTioN
cEmENTLEss moDuLAR humERAL impLANTs
mAKE
su
RE
you
ARE
usi
N
g
Th
E
DED
ic
A
TED
iN
sTR
um
ENT
s
Fo
R c
Em
ENTLE
ss mo
D
u
LAR
imp
LANT
s
Figure 80
Figure 81