AMU Powercon
Error List
A-96.250.331 / 230320
37
Error
Description
Corrective action
E001
Cond. Alarm high
– check process
– check programmed value, see
E002
Cond. Alarm low
– check process
– check programmed value, see
E007
Sample Temp. high
– check process
– check programmed value, see
E008
Sample Temp. low
– check process
E009
Sample Flow high
– check sample inlet pressure
– check programmed value, see
E010
Sample Flow low
– check sample inlet pressure
– Check flow regulating valve
– check programmed value, see
E011
Temp. shorted
– Check wiring of temperature sensor
– Check temperature sensor
E012
Temp. disconnected
– Check wiring of temperature sensor
– Check temperature sensor
E013
Case Temp. high
– check case/environment temperature
E014
Case Temp. low
– check case/environment temperature
E017
Control time-out
– Check control device or programming
in Installation, Relay contact, Relay 1/2
E018
Quality Assurance
– Perform QA Procedure using reference
instrument, e.g. AMI Inspector