AMU Powercon
Installation
A-96.250.331 / 230320
13
Example 2
SL
The Flow cell SL is used for the measurement of the acid
(cation) conductivity. Only one sensor can be used together with
the AMU Powercon. Therefore the left hand bore has to be closed
with a blind plug.
Sample
conditions
Temperature: up to 50 °C
Inlet pressure: 0.2 to 2 bar
Outlet pressure: ambient pressure
Sample flow: 5 to 20 l/h
Sample
connection
Inlet: ¼" Swagelok tube adapter (SS)
Outlet: 8/6 mm Serto tube adapter (PVDF)
Installation
The maximal recommended cable length is 15 m, select the instal
-
lation location accordingly. The flow cell must be aligned in exactly
vertical position.