56
A-96.250.331 / 230320
AMU Powercon
Program List and Explanations
5.3.2.32.4
Control Parameters
Range for each Parameter same as
.
Examples of metering devices that are pulse frequency driven are
the classic membrane pumps with a potential free triggering input.
Dosing is controlled by the repetition speed of dosing shots.
5.3.2.32.21
Pulse frequency:
Max. pulses per minute the device is able to re
-
spond to. Range: 20–300/min.
5.3.2.32.31
Control Parameters
Range for each Parameter same as
.
Dosing is controlled by the position of a motor driven mixing valve.
5.3.2.32.22
Run time:
Time needed to open a completely closed valve.
Range: 5–300 Sec.
5.3.2.32.32
Neutral zone:
Minimal response time in % of the runtime. If the re
-
quested dosing output is smaller than the response time, no
change will take place.
Range: 1–20 %
5.3.2.32.4
Control Parameters
Range for each Parameter same as
.
The relay will be activated repetitively depending on the pro
-
grammed time scheme.
5.3.2.24
Mode:
Operating mode (interval, daily, weekly)
5.3.2.340
Interval/Start time/Calendar: Dependent on options operating
mode.
5.3.2.44
Run time:
time the relay stays active.
Range: 5–32’400 Sec
5.3.2.54
Delay:
during run time plus the delay time the signal and control
outputs are held in the operating mode programmed below.
Range: 0–6’000 Sec
5.3.2.32.1
Actuator = Frequency
5.3.2.32.1
Actuator = Motor valve
5.3.2.1
Function = Timer: