A-96.250.671 / 070222
AMI Silica
Troubleshooting
68
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
E018
Reagent Pump
– shut off power
E019
FOME not connected
– shut off power
– check wiring check wiring, see
E020
FOME dirty
E022
Reagent empty
– fill reagents,
– see
Refill or replace Reagents, p. 49
E023
Sequencer
– check Sample Sequencer connection
E024
Input active
– See If Fault Yes is programmed in
Menu
E026
IC LM75
– call service
E028
Signal output open
– check wiring on signal outputs 1 and 2.
E030
I2C Rovalve
– call service
E031
Calibration Recout
– call service
E032
Wrong Frontend
– call service
E033
Sample Flow 1 low
(Sample Sequencer)
– See operating manual Sample
Sequencer
– If 2nd Sample Stream option is
installed, see E010
E034
Sample Flow 2 low
(Sample Sequencer)
– See operating manual Sample
Sequencer
– If 2nd Sample Stream option is
installed, see E010
E035
Sample Flow 3 low
(Sample Sequencer)
– See operating manual Sample
Sequencer
E036
Sample Flow 4 low
(Sample Sequencer)
– See operating manual Sample
Sequencer
Error
Description
Corrective action