A-96.250.671 / 070222
AMI Silica
Program List and Explanations
88
5.3.1.1.45
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.2
Sample Temp.:
Define at which sample temperature an alarm
should be issued.
5.3.1.2.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated.
Range: 30–70 °C
5.3.1.2.25
Alarm Low:
If the measured value rises above the alarm high value,
the alarm relay is activated.
Range: 0–20 °C
5.3.1.3
Sample Flow:
Define at which sample flow a flow alarm should be
issued.
5.3.1.3.1
Flow Alarm:
Program if the alarm relay should be activated if there is
a flow alarm. Choose between yes or no. The flow alarm will always
be indicated in the display, pending error list, saved in the message
list and the logger.
Available values: Yes or no
Note:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.3.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 100–600 B/s
5.3.1.3.35
Alarm Low:
If the measuring values falls below the programmed val
-
ue E010 will be issued.
Range: 5–80 B/s
5.3.1.4
Case Temp. high:
Set the alarm high value for temperature of elec
-
tronics housing. If the value rises above the programmed value E013
is issued.
Range: 30–75 °C
5.3.1.51
Case Temp. low:
Set the alarm low value for temperature of electron
-
ics housing. If the value falls below the programmed value E014 is
issued.
Range: -10–20 °C
5.3.2 and 5.3.3
Relay 1 and 2:
The contacts can be set as normally open or normal
-
. The function of re
-
lay contacts 1 or 2 are defined by the user.