A-96.250.671 / 070222
AMI Silica
Program List and Explanations
94
5.3.4.4
Fault:
5.3.4.5
Delay:
Time which the instrument waits, after the input is deactivat
-
ed, before returning to normal operation.
Range: 0–6‘000 Sec
5.4 Miscellaneous
5.4.1
Language:
Set the desired language.
5.4.2
Set defaults:
Reset the instrument to factory default values in three
different ways:
Calibration
: Sets calibration values back to default. All other
values are kept in memory.
In parts
: Communication parameters are kept in memory. All other
values are set back to default values.
Completely
: Sets back all values including communication
parameters.
5.4.3
Load Firmware:
Firmware updates should be done by instructed ser
-
vice personnel only.
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes:
Message E024 is issued and stored in the message
list. The Alarm relay closes when input is active.
Language
German
English
French
Spanish
Set defaults
no
Calibration
In parts
Completely
Load Firmware
no
yes