A-96.250.671 / 070222
AMI Silica
Program List and Explanations
95
5.4.4
Password:
Select a password different from 0000 to prevent
unauthorized access to the following menus:
5.4.4.1
Messages
5.4.4.2
Maintenance
5.4.4.3
Operation
5.4.4.4
Installation.
Each menu may be protected by a
different
password.
If you forgot the passwords, contact the closest SWAN representative.
5.4.5
Sample ID:
Identify the process value with any meaning full text,
such as KKS number.
5.4.6
Line Break Detection:
If activated, error message E028 is shown in
case of line break on signal outputs 1 and 2.
5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
5.5.1
Protocol:
Profibus
5.5.20
Device address:
Range: 0–126
5.5.30
ID-Nr.:
Range: Analyzer; Manufacturer; Multivariable
5.5.40
Local operation:
Range: Enabled, Disabled
5.5.1
Protocol:
Modbus RTU
5.5.21
Device address:
Range: 0–126
5.5.31
Baud Rate:
Range: 1200–115 200 Baud
5.5.41
Parity:
Range: none, even, odd
5.5.1
Protocol:
USB stick
Only visible if an USB interface is installed. No further settings are
possible.
5.5.1
Protocol:
HART
5.5.24
Device address:
Range: 0–63