A-96.250.671 / 070222
AMI Silica
Program List and Explanations
78
2.2.1.5
Ver. History:
Review verifications values of the last verifications:
Absorbance:
Measured absorbance of the reference kit.
Reference value
: True value of the reference kit according to label.
2.2.2
Miscellaneous
:
2.2.2.1
Case Temp:
Shows the actual temperature in [°C] inside the trans
-
mitter.
State Machine
: Number 1–7 are assigned to the initial start up pro
-
cedure.
Number 8–23 are assigned to the measuring cycle.
Number 24 is assigned to the manual filling function.
Number 25–31 are assigned to the calibration procedure.
Number 32–35 are assigned to the verifying procedure.
2.3 Sample
2.3.1
Sample ID
:
Review the programmed code. The code is defined by
the user to identify the sample point in the plant.
Temperature:
Shows the Actual temperature in °C.
Sample flow:
Shows the actual sample flow [B/s] (bubbles per sec
-
ond). Sample flow must be above 5 B/s.
Machine state 6-way valve position
peristaltic pump
9
5
on
10
5
on
11
5
off
12
1
on
13
2
on
14
5
on
15
5
off
16
3
on
17
5
on
18
5
off
19
4
on
20
5
on
21
5
off
22
23