6.7. Cleaning the Photometer
Clean the photometer after indication by alarm (E020, FOME dirty).
Before opening the photometer, switch off the instrument according
Stop of Operation for Maintenance, p. 48
Material
Small brush.
Procedure
1
Close the flow regulating valve [A].
2
Wait until the sample flow through the photometer has stopped
3
Unscrew the cover [B] from the photometer [C].
4
Clean the Photometer with a small brush [D].
5
Screw the cover to the photometer.
6
Open the flow regulating valve.
A
B
C
D
Flow regulating valve
Photometer cover
Photometer
Brush