A-96.250.671 / 070222
AMI Silica
Maintenance
62
6.10. Fill or Flush Reagent System
Fill or flush the reagent tubing:
after refilling the reagent containers,
before a system shut-down to flush the system with demin
water until no more reagent is left in the system.
6.11. Longer Stop of Operation
1
Stop of Operation for Maintenance,
2
Empty the measuring cell of the photometer, for example with a
pipette and dry it with a soft tissue.
3
Relax the occlusion frame of the Peristaltic pump. See
.
Press 4 x [Exit]
3.2.2
Fill System
Verification
Service
Manual filling
Enter
3.2.2.5
Fill System
Progress
<Enter> to stop
3.4.5
Fill System
Progress
Done