P-97 FLAMING/BROWN MICROPIPETTE PULLER OPERATION MANUAL – REV. 2.43 - DOM (20161118)
42
5.2.2
Problem: Display blank, fan not on, or locked up/overlap.
1.
Check power cord and wall a.c. outlet.
2.
If the unit still does not work after verifying it is properly plugged in, remove the power
cord and check the fuse. If the fuse has blown, a failure in components that are not
serviceable by the user has likely occurred. Contact Sutter Instrument Company
Technical Support.
3.
If the fuse is still good, the unit is properly plugged in, and it still does not work, a failure
in components that are not serviceable by the user has likely occurred. Contact Sutter
Instrument Company Technical Support.
5.2.3
Problem: Display only shows a cursor.
1.
The microprocessor has failed to properly initialize the display. Press RESET and the
display should show the proper power-up message. Always allow at least 5 seconds before
turning the power back on. If the display still shows a single cursor, contact Sutter
Instrument Company Technical Support.
5.2.4
Problem: Displayed program values are not correct
1.
Make sure that values were not changed by another user. Always write down the program
values and the ramp-test value and keep them in a secure place.
2.
If the values entered are not held when the power is turned off, a the embedded battery
in the SRAM memory chip has probably failed. Contact Sutter Instrument Company
Technical Support.
5.3
Technical Support
For further assistance, contact Sutter Instrument Technical Support at:
(415) 883-0128 or [email protected]